Clean dome internal-projector window and check Mylar ring mounting — 1.34
Clean the dome cover's red transparent Mylar sheet and confirm the Mylar ring is mounted flush against the dome and front covers, so no gap is left for fluids to seep into the system (periodic maintenance item 1.34).
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Hi-clean (washing liquid) | n/a in source | |
| Rag | n/a in source |
Safety
Hazard: Mylar ring not mounted securely against the dome cover; source labels this call-out "CAUTION" — reclassified NOTICE here because the stated outcome is system damage/image-quality deterioration, not personal injury
Consequence: Fluids such as blood, contrast medium, etc. may leak into the system through the gap between the Mylar sheet and the cover, resulting in damage to the system or deterioration of image quality
Avoidance: Mount the Mylar ring securely, so it is in contact with both the dome cover and the front cover with no floating or cambered sections
Full step-by-step for Clean dome internal-projector window and check Mylar ring mounting — 1.34
All 5 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry guides
Source: 2Y201-061E*E p.74-75. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.