Gantry terminal-board looseness, screw/nut, cable and final cleaning checks — 1.55-1.58
Confirm the Aquilion 16 gantry's 200-V input terminal boards (TB1, TB2) and their screws/nuts are not loose, visually inspect the gantry and couch cables, and clean the entire gantry (including dust on the PWB) — periodic gantry inspection items 1.55 through 1.58, the closing items of the Gantry section of this maintenance volume.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Flat-bladed parallel-tip screwdriver | n/a in source |
Safety
Hazard: Performing the checks in items 1.55, 1.56, or 1.57 (the gantry 200-V input terminal boards, their screws/nuts, and the gantry/couch cables) while any system power supply is still ON.
Consequence: Source states only that these checks must be performed with all system power supplies turned OFF; it does not restate a separate consequence phrase for this specific note. The general electric-shock/injury hazard from live 200-V terminal boards (TB1, TB2) and connected cables applies — see this volume's general Safety Precautions (WARNING 2 / DANGER 4, Related) for the equivalent hazard stated in full.
Avoidance: Turn OFF all system power supplies before performing items 1.55, 1.56, and 1.57. Source's DANGER note scopes explicitly to these three items; it does not name item 1.58 (cleaning) — see the note under Steps below.
Full step-by-step for Gantry terminal-board looseness, screw/nut, cable and final cleaning checks — 1.55-1.58
All 8 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry guides
Source: 2Y201-061E*E p.97-99. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.