Troubleshoot: couch will not go up/down and oil leakage
Use this fault-isolation flow when the Aquilion 16 patient couch's vertical (lift) drive won't raise, won't lower, or shows hydraulic oil leakage — the source manual treats all three symptoms as one linked flowchart across two pages.
Applies to
Aquilion 16 (TSX-101A) — Patient Couch Mechanism volume, source 2D201-062
Personnel
1 technician · Intermediate — hydraulic + electrical fault isolation on the couch vertical (lift) drive; mixes a live-circuit check at mains level (source calls for >=180 VAC, and a rated-voltage check at the solenoid valve) with mechanical observation (pump noise, oil level, sinking rate), and ends by handing off to dedicated replacement procedures for the tilt encoder / solenoid valve / hydraulic unit (skill assessment derived from procedure content — not an OEM-stated skill level)
Steps
9 steps
Source
2D201-062E*F p.32-33
Full step-by-step for Troubleshoot: couch will not go up/down and oil leakage
All 9 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Table (Vertical) guides
Adjust the vertical (lowering) motion speed5 stepsChange the vertical stop positions (height photosensor adjustment)2 stepsJack Up the Patient Couch When It Will Not Rise From the DOWN Limit4 stepsLower the Patient Couch from a Raised Position After a Failure (200 VAC Jig)4 stepsReplace the hydraulic unit (cylinder, pump, oil tank)25 stepsRoute the internal cables at the lifting linkage8 steps
View all Aquilion 16 (TSX-101A) guidesSource: 2D201-062E*F p.32-33. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.