Troubleshoot: couch will not go up/down and oil leakage

Use this fault-isolation flow when the Aquilion 16 patient couch's vertical (lift) drive won't raise, won't lower, or shows hydraulic oil leakage — the source manual treats all three symptoms as one linked flowchart across two pages.

Applies to
Aquilion 16 (TSX-101A) — Patient Couch Mechanism volume, source 2D201-062
Personnel
1 technician · Intermediate — hydraulic + electrical fault isolation on the couch vertical (lift) drive; mixes a live-circuit check at mains level (source calls for >=180 VAC, and a rated-voltage check at the solenoid valve) with mechanical observation (pump noise, oil level, sinking rate), and ends by handing off to dedicated replacement procedures for the tilt encoder / solenoid valve / hydraulic unit (skill assessment derived from procedure content — not an OEM-stated skill level)
Steps
9 steps
Source
2D201-062E*F p.32-33

Full step-by-step for Troubleshoot: couch will not go up/down and oil leakage

All 9 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Couch/Table (Vertical) guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-062E*F p.32-33. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.