Replace the hydraulic unit (cylinder, pump, oil tank)
Replace the Aquilion 16 patient couch's vertical-drive hydraulic unit — cylinder, motor pump, and oil tank — as a single assembly, when the height (vertical) drive malfunctions.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Straps | n/a in source | 2 × 3 m |
| Support bars | n/a in source | 2 × 50 mm × 50 mm × 600 mm |
| Phillips screwdriver | n/a in source | No. 2, medium size |
| Hexagonal wrenches | n/a in source | 3 mm and 5 mm |
| Hammer | n/a in source | — |
| Punch | n/a in source | — |
| Screw-locking compound | n/a in source | — |
| Wire nipper | n/a in source | — |
| File | n/a in source | — |
| Hydraulic unit, factory-supplied replacement | n/a in source | 1 |
Safety
Hazard: Hydraulic hose bent tighter than its minimum permissible bending radius (45 mm) while being handled during mounting.
Consequence: The hose is permanently damaged and can no longer be used.
Avoidance: Be extremely careful when handling the hydraulic hose; do not bend it tighter than the 45 mm minimum permissible bending radius.
Full step-by-step for Replace the hydraulic unit (cylinder, pump, oil tank)
All 25 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Table (Vertical) guides
Source: 2D201-062E*F p.55-61. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.