ProcedureCouch/Table (Vertical)Adjust

Adjust the vertical (lowering) motion speed

This procedure adjusts how fast the Aquilion 16 patient couch lowers, by matching the openings of the two solenoid throttle valves — L21 and L24 — that control descent speed.

Applies to
Aquilion 16 (TSX-101A) — Patient Couch Mechanism volume, source 2D201-062, section 6.1 (Adjustments and Cleaning)
Personnel
1 technician · Basic-to-intermediate — hand adjustment of two solenoid throttle-valve knobs and their shared locking nut (fitted to L21 only); no electrical measurement or component replacement is described for this section (skill assessment derived from procedure content — not an OEM-stated skill level)
Steps
5 steps
Source
2D201-062E*F p.62

Safety

CAUTION

Hazard: Being caught between the arm or upper frame while adjusting the lowering speed, or while working on parts inside the lifter arms.

Consequence: Risk of being crushed by the arm or upper frame.

Avoidance: Be careful not to be crushed by the arm or upper frame when adjusting the lowering speed or working on the parts inside the lifter arms.

NOTICE

Hazard: Leaving the adjustment knob detached from the solenoid valve while the valve is open.

Consequence: Oil leaks from the knob hole.

Avoidance: Never leave the knob detached from the valve. After adjusting the speed, be sure to tighten the knob's locking nut (turn the nut clockwise to tighten).

Full step-by-step for Adjust the vertical (lowering) motion speed