Adjust the vertical (lowering) motion speed
This procedure adjusts how fast the Aquilion 16 patient couch lowers, by matching the openings of the two solenoid throttle valves — L21 and L24 — that control descent speed.
Safety
Hazard: Being caught between the arm or upper frame while adjusting the lowering speed, or while working on parts inside the lifter arms.
Consequence: Risk of being crushed by the arm or upper frame.
Avoidance: Be careful not to be crushed by the arm or upper frame when adjusting the lowering speed or working on the parts inside the lifter arms.
Hazard: Leaving the adjustment knob detached from the solenoid valve while the valve is open.
Consequence: Oil leaks from the knob hole.
Avoidance: Never leave the knob detached from the valve. After adjusting the speed, be sure to tighten the knob's locking nut (turn the nut clockwise to tighten).