Jack Up the Patient Couch When It Will Not Rise From the DOWN Limit

When the patient couch is disabled and cannot move up from the DOWN limit under its own power, this procedure raises it with an external automotive jack so the parts inside the couch can be checked.

Applies to
Aquilion 16 (TSX-101A) — patient couch CBTB-016A / CBTB-016B, vertical (height) lift mechanism
Personnel
1 technician · Basic/Intermediate — cover removal (front under cover + bellows) and manual operation of an automotive jack; no electrical work and no specialized service tooling called out in source for this procedure (skill assessment derived from procedure content — not an OEM-stated skill level)
Steps
4 steps
Source
2D201-062E*F p.37-38

Tools & parts

ItemPart numberQty
Automotive jack, rated for at least 500 kgn/a in source
Platform, 9 cm × 9 cmn/a in source
Square piece of woodn/a in source

Safety

CAUTION

Hazard: Performing the internal inspection with the patient couch held up by an automotive jack, if the jack slips.

Consequence: The upper frame may fall, resulting in injury.

Avoidance: Place a square piece of wood between the frame and the floor before starting the inspection, so that safety is preserved even if the jack slips.

Full step-by-step for Jack Up the Patient Couch When It Will Not Rise From the DOWN Limit

All 4 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Couch/Table (Vertical) guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-062E*F p.37-38. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.