Jack Up the Patient Couch When It Will Not Rise From the DOWN Limit
When the patient couch is disabled and cannot move up from the DOWN limit under its own power, this procedure raises it with an external automotive jack so the parts inside the couch can be checked.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Automotive jack, rated for at least 500 kg | n/a in source | |
| Platform, 9 cm × 9 cm | n/a in source | |
| Square piece of wood | n/a in source |
Safety
Hazard: Performing the internal inspection with the patient couch held up by an automotive jack, if the jack slips.
Consequence: The upper frame may fall, resulting in injury.
Avoidance: Place a square piece of wood between the frame and the floor before starting the inspection, so that safety is preserved even if the jack slips.
Full step-by-step for Jack Up the Patient Couch When It Will Not Rise From the DOWN Limit
All 4 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Table (Vertical) guides
Source: 2D201-062E*F p.37-38. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.