Gantry rotation speed check and fine adjustment
Check the Aquilion 16 gantry's rotation speed over a KGTSM terminal session, then — since Aquilion performs no automatic speed correction — use DIP switch DSW12 to finely correct that speed back to within ±3% of standard whenever a check shows it out of tolerance.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| PC | n/a in source | |
| Service cable (GR-1) | GR-1 (standard accessory) | |
| Terminal program (e.g. Hyper Terminal) | n/a in source |
Safety
Hazard: Rotating the gantry as part of this speed-check and fine-adjustment procedure.
Consequence: Not explicitly stated in source beyond the instruction itself.
Avoidance: Be careful when rotating the gantry throughout this procedure (source callout, p.108, printed at the start of subsection 7.4.1 Outline).
Hazard: Setting the KGTSM panel switch (SW1) to OFF (upward) before all maintenance-panel operations used in this procedure are complete, or not resetting the KGTSM after using the panel.
Consequence: Not explicitly stated in source beyond the instruction itself.
Avoidance: Do not set the panel switch (SW1) to OFF (upward) before all operations from the maintenance panel are completed. After using the maintenance panel, be sure to set the panel switch OFF and reset the KGTSM (source callout, p.109).
Full step-by-step for Gantry rotation speed check and fine adjustment
All 7 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry/Rotation guides
Source: 2D201-063E*G p.107-111. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.