Pular para o conteúdo principal
ProcedureGantry/RotationReplace

Replace the rotation-direction photo sensor unit

Replace the Aquilion 16 gantry's rotation-direction photo sensor unit behind the MUDAT external cover, reusing its mounting spacer, then confirm the gantry's rotation speed still checks out within tolerance.

Applies to
Aquilion 16 (TSX-101A)
Personnel
1 technician · Intermediate — mechanical removal/replacement of the rotation-direction photo sensor behind the MUDAT external cover (cover access, 2 mounting screws for the cover, 2 for the sensor, plus a spacer and a shim that must be handled carefully and reused), followed by a rotation-speed check — and, if that check fails, the cross-referenced fine-adjustment or MR-sensor-adjustment procedure — to confirm the replacement didn't disturb rotation timing (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
6 steps
Source
2D201-063E*G p.152-153

Tools & parts

ItemPart numberQty
Photo sensor unit (replacement)n/a in source
Spacern/a in source
Shimn/a in source
MUDAT external cover mounting screwsn/a in source
Photo sensor unit mounting screwsn/a in source

Full step-by-step for Replace the rotation-direction photo sensor unit

All 6 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Gantry/Rotation guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-063E*G p.152-153. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.