ProcedureGantry/RotationReplace
Replace the rotation-direction photo sensor unit
Replace the Aquilion 16 gantry's rotation-direction photo sensor unit behind the MUDAT external cover, reusing its mounting spacer, then confirm the gantry's rotation speed still checks out within tolerance.
Applies to
Aquilion 16 (TSX-101A)
Personnel
1 technician · Intermediate — mechanical removal/replacement of the rotation-direction photo sensor behind the MUDAT external cover (cover access, 2 mounting screws for the cover, 2 for the sensor, plus a spacer and a shim that must be handled carefully and reused), followed by a rotation-speed check — and, if that check fails, the cross-referenced fine-adjustment or MR-sensor-adjustment procedure — to confirm the replacement didn't disturb rotation timing (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
6 steps
Source
2D201-063E*G p.152-153
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Photo sensor unit (replacement) | n/a in source | |
| Spacer | n/a in source | |
| Shim | n/a in source | |
| MUDAT external cover mounting screws | n/a in source | |
| Photo sensor unit mounting screws | n/a in source |
Full step-by-step for Replace the rotation-direction photo sensor unit
All 6 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry/Rotation guides
Adjust the MR sensor and MR sensor PWB20 stepsReplace the MR sensor and MR sensor PWB12 stepsGantry rotation interlock when servicing with cover open11 stepsGantry rotation speed check and fine adjustment7 stepsGantry rotation system operation and troubleshooting (S/N Axxxxx)0 stepsGantry rotation system operation and troubleshooting (S/N Bxxxxx)5 steps