C0030003 — Raw-data disk fault on the CT recon disk array (RDD)
Canon Medical (Toshiba) CT error code, reported by the RTM RDD (raw data disk array).
What C0030003 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code C0030003 indicates: Raw-data disk fault on the CT recon disk array (RDD). The code originates in the RTM RDD — raw data disk array section of the RTM RDD (raw data disk array). Severity level A1 (Alarm — high severity).
Reported action: Stop scanning and check the RDD status LEDs (power, fan, HDD) on the console's disk array. Note the exam that produced the alarm and escalate to your service representative: a fault in the raw data disk array puts the raw data used for reconstruction at risk, and recovery means replacing the affected disk carrier and reconfiguring the array with parity..
Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 102:Device RTM RDD group of the console error table, where the OEM text reads "Error in RDD.", graded A1 (high-severity alarm). RDD is the Raw Data Disk Array that holds the scan raw data used for reconstruction; on the Aquilion 16 console it is the fibre-channel disk array mounted in the REC BOX.
Full troubleshooting for C0030003
The complete diagnosis, cause tree, and step-by-step remedy for the RTM RDD — raw data disk array subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated RTM RDD — raw data disk array codes