C0030001 — Raw data disk array (RDD) error on the CT console
Canon Medical (Toshiba) CT error code, reported by the RTM RDD (raw data disk array).
What C0030001 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code C0030001 indicates: Raw data disk array (RDD) error on the CT console. The code originates in the RTM RDD — raw data disk array section of the RTM RDD (raw data disk array). Severity level A1 (Alarm — high severity).
Reported action: Check the disk array before scanning again: on the Aquilion 16 console the RDD sits in the REC BOX and its LEDs report power, fan and HDD status, so a failed drive or fan is usually visible on the front panel. Note which scan raised the alarm and have your service representative inspect the array — replacing a disk also requires reconfiguring the RDD with parity, which is service work..
Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 102:Device RTM RDD group of the console error table, where the OEM text reads "Error in RDD.", graded A1 (high-severity alarm). RDD is the Raw Data Disk Array that holds the scan raw data used for reconstruction; on the Aquilion 16 console it is the fibre-channel disk array mounted in the REC BOX.
Full troubleshooting for C0030001
The complete diagnosis, cause tree, and step-by-step remedy for the RTM RDD — raw data disk array subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated RTM RDD — raw data disk array codes
Related service procedures
Source: errmsg.eng — 102:Device RTM RDD table. Reference information re-authored by FSELIB — not a copy of the OEM manual.