Pular para o conteúdo principal
Information / dialogSYNC — respiratory gating

B8320000Respiratory-gating signal abnormal — respiratory sync signal error (5102)

Canon Medical (Toshiba) CT error code, reported by the SYNC device — respiratory gating.

What B8320000 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code B8320000 indicates: Respiratory-gating signal abnormal — respiratory sync signal error (5102). The code originates in the SYNC — respiratory gating section of the SYNC device — respiratory gating. Severity level A1 (Information / dialog).

Reported action: The signal is already bad before the scan is confirmed, so check the respiratory gating system and its connection as the message instructs: sensor/belt placement and tension on the patient, the sensor cable and its connector at the SYNC interface, and a visible respiratory waveform on the monitor. Re-seat the sensor and confirm a stable trace before pressing confirm; if the waveform is good and the code persists, follow the SYNC device checks in the service manual and call service..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003, Aquilion ONE (V4.2), Aquilion S16, Asteion Super4 and Aquilion 16 CT systems, on the SYNC device that handles the respiratory-gating signal. The service table pairs this code with detail 5102: a respiratory sync signal error, with the gating signal already abnormal before the confirm button is pressed. The same short message text is also published under code B8F20000 (detail 5103).

Full troubleshooting for B8320000

The complete diagnosis, cause tree, and step-by-step remedy for the SYNC — respiratory gating subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related SYNC — respiratory gating codes

Related service procedures

Discuss B8320000 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng). Reference information re-authored by FSELIB — not a copy of the OEM manual.