Pular para o conteúdo principal
Information / dialogKeyboard (Nabipad)

B4100000Keyboard error reported by the Nabipad device

Canon Medical (Toshiba) CT error code, reported by the Nabipad (keyboard).

What B4100000 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code B4100000 indicates: Keyboard error reported by the Nabipad device. The code originates in the Keyboard (Nabipad) section of the Nabipad (keyboard). Severity level 1 (Information / dialog).

Reported action: Confirm the keyboard is properly connected and that no key is stuck or being held down, then restart the workstation and try again. If the keyboard error returns, the Nabipad unit or its interface has to be checked by service — note whether specific keys are affected or the whole device is dead..

Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 6:Device Nabipad table, where the OEM text reads "Error in KeyBoard."

Full troubleshooting for B4100000

The complete diagnosis, cause tree, and step-by-step remedy for the Keyboard (Nabipad) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Keyboard (Nabipad) codes

Related service procedures

Discuss B4100000 with other field engineers in the FSELIB forum

Source: errmsg.eng — 6:Device Nabipad table. Reference information re-authored by FSELIB — not a copy of the OEM manual.