Pular para o conteúdo principal
Alarm — high severityRDD (Disk Array)

80020002RDD (disk array) error — parameter set rejected

Canon Medical (Toshiba) CT error code, reported by the RDD (Disk Array).

What 80020002 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 80020002 indicates: RDD (disk array) error — parameter set rejected. The code originates in the RDD (Disk Array) section of the RDD (Disk Array). Severity level A1 (Alarm — high severity).

Reported action: Note the scan or transfer running when the alarm was raised and retry it. Check the RDD status and the system log, and call service if the disk array keeps rejecting the parameter set..

Listed for Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003, under the 40:Device RDD table, where the OEM text reads "Error in RDD (parameter set)." It is classified as a high-severity alarm (A1). On the Aquilion 16, the RDD is the Raw Data Disk Array that the preprocessing/data transmission unit writes the raw scan data to.

Full troubleshooting for 80020002

The complete diagnosis, cause tree, and step-by-step remedy for the RDD (Disk Array) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RDD (Disk Array) codes

Related service procedures

Discuss 80020002 with other field engineers in the FSELIB forum

Source: errmsg.eng — 40:Device RDD table. Reference information re-authored by FSELIB — not a copy of the OEM manual.