Pular para o conteúdo principal
Information / dialogReconstruction

0000002DAbnormal RGR data found in a reconstruction request

Canon Medical (Toshiba) CT error code, reported by the APPL — RGR (respiratory-gated reconstruction) data validation.

What 0000002D means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code 0000002D indicates: Abnormal RGR data found in a reconstruction request. The code originates in the Reconstruction section of the APPL — RGR (respiratory-gated reconstruction) data validation. Severity level 0 (Information / dialog).

Reported action: Delete the affected request from the reconstruction queue and retry reconstruction using the conventional method; check the respiratory-synchronization waveform in the raw data before attempting respiratory-gated reconstruction again..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003 and Aquilion ONE (V4.2) CT systems — the same OEM event-log table uses 'RGR' elsewhere for respiratory-gated reconstruction. This code is raised when abnormal RGR data is found; the OEM message instructs deleting the request from the reconstruction queue and retrying with the conventional method, and the parallel Japanese-language OEM text for this code specifically names the respiratory-synchronization waveform in the raw data.

Full troubleshooting for 0000002D

The complete diagnosis, cause tree, and step-by-step remedy for the Reconstruction subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Reconstruction codes

Related service procedures

Discuss 0000002D with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng, Aquilion RX V4-62R004 / Aquilion ONE V4.2); cross-referenced with the OEM EMS event-log code table (EMS_Errcode.ini, EMS V3.20.3), including its Japanese-language string, for the same code. Reference information re-authored by FSELIB — not a copy of the OEM manual.