00000014 — Image data disk battery error
Canon Medical (Toshiba) CT error code, reported by the Image data disk unit.
What 00000014 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 00000014 indicates: Image data disk battery error. The code originates in the Image data disk unit section of the Image data disk unit. Severity level 1 (Information / dialog).
Reported action: Check the image data disk unit's backup battery and replace it if faulted or due per the maintenance schedule. If what actually showed was only "Couch is free now.", that is the unrelated benign status message and needs no service action..
Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003 (the disk battery text specifically on Aquilion RX/V4-62R004, V3.20R003 and Aquilion ONE/V4.2). This code number is shared across the 1:SS Display Message and 80:Device APPL tables in errmsg.eng, whose OEM texts are "Couch is free now." and "Error in Image data disk (battery)." — the image data disk battery fault is what matters; "Couch is free now." is the routine status use of the same code number.
Full troubleshooting for 00000014
The complete diagnosis, cause tree, and step-by-step remedy for the Image data disk unit subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Image data disk unit codes
Related service procedures
Source: errmsg.eng — 1:SS Display Message / 80:Device APPL tables. Reference information re-authored by FSELIB — not a copy of the OEM manual.