Information / dialogECG gating

0000000FAbnormal ECG data found

Canon Medical (Toshiba) CT error code, reported by the APPL (application software).

What 0000000F means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code 0000000F indicates: Abnormal ECG data found. The code originates in the ECG gating section of the APPL (application software). Severity level 0 (Information / dialog).

Reported action: The system found abnormal ECG data — the ECG trace that ECG-gated scanning and reconstruction rely on is not usable as recorded. Check the ECG electrodes, cable and signal quality on the patient and repeat the acquisition; if the recorded ECG cannot be used, process the study without ECG gating..

Listed only in the 80:Device APPL table, for Aquilion RX/V4-62R004, V3.20R003 and Aquilion ONE/V4.2. The OEM text is "Abnormal ECG data was found."