0000000D — Failed to transfer the scan condition
Canon Medical (Toshiba) CT error code, reported by the APPL (application software).
What 0000000D means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code 0000000D indicates: Failed to transfer the scan condition. The code originates in the Scan operation section of the APPL (application software). Severity level 0 (Information / dialog).
Reported action: The system failed to transfer a scan condition (the parameter set for the scan). Re-select or re-enter the scan protocol and retry; if the transfer keeps failing, restart the exam plan and, if it still fails, restart the system..
Listed only in the 80:Device APPL table, for Aquilion RX/V4-62R004, V3.20R003 and Aquilion ONE/V4.2. The OEM text is "Failed to transfer a scan condition."
Full troubleshooting for 0000000D
The complete diagnosis, cause tree, and step-by-step remedy for the Scan operation subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Scan operation codes
Related service procedures
Source: errmsg.eng — 80:Device APPL table. Reference information re-authored by FSELIB — not a copy of the OEM manual.