Canon Medical (Toshiba) CT error code, reported by the APPL (application software).

What 0000000D means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 CT systems, status code 0000000D indicates: Failed to transfer the scan condition. The code originates in the Scan operation section of the APPL (application software). Severity level 0 (Information / dialog).

Reported action: The system failed to transfer a scan condition (the parameter set for the scan). Re-select or re-enter the scan protocol and retry; if the transfer keeps failing, restart the exam plan and, if it still fails, restart the system..

Listed only in the 80:Device APPL table, for Aquilion RX/V4-62R004, V3.20R003 and Aquilion ONE/V4.2. The OEM text is "Failed to transfer a scan condition."

Full troubleshooting for 0000000D