InstallationSoftware / InstallationInstallationInstall

Initialize the system: model/tube setup, special config, RDD/RAID, host name

Once the system software is on the disk, the freshly installed Aquilion 16 still doesn't know its own name, its tube, its site's network address, or any of the dozen option toggles a real install needs — this section walks it through all of that at the Service Engineer's level, configures the RDD as a parity RAID, gives it a host name and IP, and closes by handing the system back its own site-specific identity from the pre-install MO backup.

Applies to
Aquilion 16 (TSX-101A) — system software volume 2D260-074E*B, source p.155-169, covering source Chapter 9 sections 9.2.4 "Initializing the system" (System/Model Name, Tube & HFG, exam-plan/calibration initialization, Change Language, Special Configuration, RDD Setup, Change Computer Name/IP), 9.2.5 "Operational check," and 9.2.6 "Restoring the site-specific information." Source runs this sequence the same way for the scan (image acquisition) and image processing systems, with one explicit exception: RDD Setup (Step Group 7 below) is "not required for the image processing system." Source states no other model- or workstation-variant split for this section.
Personnel
Basic — a console menu/GUI procedure (Change System → Service Engineer's → password; pull-down selections for name/tube/language; Special Configuration checkboxes and menus; RDD Setup buttons; network-field entry) with no board-level work or physical tooling described anywhere in this section. The one item that calls for care rather than technique is the CBTB couch-type selection inside Special Configuration, which source itself flags with a CAUTION because it is easy to set from a menu without checking the physical couch in front of you (see Safety). (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
25 steps
Source
2D260-074E*B p.155-169

Safety

CAUTION

Hazard: The couch type selected under Special Configuration's CBTB Option (CBTB-014A / CBTB-016A / CBTB-016B) does not match the couch physically installed on the system.

Consequence: Scanning may not be performed as planned (source's own wording).

Avoidance: Confirm the actual patient couch model before touching the CBTB Option field, and set it to match exactly.

Full step-by-step for Initialize the system: model/tube setup, special config, RDD/RAID, host name

All 25 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Software / Installation guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D260-074E*B p.155-169. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.