Couch hydraulic leakage, lubrication, tightening, natural-fall, DC power and sensor checks — 2.12-2.19
Run this eight-item block of the patient-couch periodic inspection table — hydraulic-circuit oil-leak check, cylinder-rod lubrication, cable check, tightening check, anchor-bolt check, natural-falling-rate measurement, DC power supply check, and sensor check (items 2.12 through 2.19) — to catch hydraulic and mechanical wear on the couch's lift system before it becomes a "won't hold position" or "won't move" failure.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Lubricant (item 2.13) | **Rust Jet (KOHZAI)** or **Power Green (BALBIS)** | |
| Scale, or an adding tape (item 2.17) | n/a in source (no brand/model given) |
Safety
Hazard: Performing the sensor check (item 2.19) — which requires the couch lower cover removed to reach the SLIDE and HEIGHT sensors — without first isolating the gantry breaker NFB1.
Consequence: Source states this is to ensure safety (no separate injury named for this item).
Avoidance: Turn OFF the gantry breaker NFB1 before performing the sensor check with the couch lower cover removed.
Full step-by-step for Couch hydraulic leakage, lubrication, tightening, natural-fall, DC power and sensor checks — 2.12-2.19
All 1 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Table guides
Source: 2Y201-061E*E p.106-108. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.