Partially disassemble the patient couch for carrying-in (narrow route)

Partially disassemble the patient couch — upper covers, couch top, handle, belt, couch-top slide frame, LM guide, and rear-end frame — to shorten its carrying-in length for routes too narrow for the fully assembled couch.

Applies to
Aquilion 16 (TSX-101A) — CBTB-016 configuration only (source appendix heading: "For CBTB-016 Only")
Personnel
1 technician · Intermediate — sequential removal of several M8-bolted mechanical assemblies (belt fixing plate, couch-top slide frame, LM guide, rear-end frame), building on the cover/couch-top/handle removal in subsection 2.7, plus disconnecting connectors and grounding wires from the rear-end frame and reinstalling to a specified torque (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
5 steps
Source
2D201-062E*F p.79-82

Tools & parts

ItemPart numberQty
Upper covers, couch top, handlen/a in source— (see subsection 2.7)
Belt + belt fixing platen/a in source4 × M8 bolts
Couch-top slide framen/a in source6 × M8 bolts
LM (linear) guiden/a in source17 × M8 bolts
Rear-end framen/a in source7 × M8 bolts