Reset the overcurrent relay (horizontal and vertical motors)
The Aquilion 16 patient couch protects both its horizontal (longitudinal/slide) motor and its vertical (lift/hydraulic pump) motor with an overcurrent relay that trips to stop the motor from smoking or burning — this atom covers how to reset each one, and what you must check first.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Jig cover (between gantry and couch) | n/a in source |
Safety
Hazard: Resetting either the horizontal- or vertical-movement-motor overcurrent relay without first checking whether the circuit is shorted or the drive unit is locked (a foreign substance caught in it, etc.).
Consequence: The overcurrent relay exists specifically to stop the motor from smoking and burning when an overcurrent flows; resetting it without finding and correcting the cause of the overcurrent is dangerous, because the same fault can trip the relay again and let the motor smoke or burn.
Avoidance: Before resetting either overcurrent relay, check whether the circuit is shorted or the drive unit is locked (foreign object caught) and correct the cause. Do not reset the relay without determining and correcting the cause of the overcurrent.
Full step-by-step for Reset the overcurrent relay (horizontal and vertical motors)
All 6 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Table guides
Source: 2D201-062E*F p.75-76. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.