Connect warning lamp, door switch and inter-unit cables
Wire the warning lamp and door switch into TB2/TB3 with the correct fuse and JP1 jumper mode, then land the patient-couch, console (CPU BOX/REC BOX) and gantry-side inter-unit cables at their labeled connectors and clamp points, completing the electrical hookup of the Aquilion 16 before the pre-power checks begin.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| 250 V, 0.5 A fuse | n/a in source | 2 |
| KGTSP PWB JP1 jumper | n/a in source — full 4-position table (incl. pins 7-8, unused here) in [OPCONTA, GCIFA and KGTSP PWB settings and connectors](toshiba-ct-aquilion16-control-opconta-gcifa-kgtsp-pwb.md) | 1, set to one of three positions |
| Jig cover mounting bracket | n/a in source | 1 |
| FCP40 (POWER) | CX77-96088-1 *(§8.1 cable-designator legend, p.67 — not restated on p.73-80)* | 1 |
| FCS46 (SIGNAL) | CX77-80261-2 *(§8.1 legend, p.67)* | 1 |
| FCS12 (SIGNAL) | CX77-96087-1 *(§8.1 legend, p.67)* | 1 |
| FCE01 (EARTH) | CX77-96089-1 *(§8.1 legend, p.67)* | 1 |
| S01 | BSX70-0391 *(§8.1 legend, p.67)* | 1 |
| S02 | CX70-02675-1 *(§8.1 legend, p.67)* | 1 |
| S03 | CX70-02735-1 *(§8.1 legend, p.67)* | 1 |
| P01 | CX70-03024-1 *(§8.1 legend, p.67)* | 1 |
| P04 | CX70-02725-1 *(§8.1 legend, p.67)* | 1 |
| P02 | PX70-02582-1 *(§8.1 legend, p.67)* | 1 |
| P03 | PX70-02583-1 *(§8.1 legend, p.67)* | 1 |
| E01 | PX70-02585-1 *(§8.1 legend, p.67)* | 1 |
| CS01 | BSX74-1358-01 *(§8.1 legend, p.67)* | 1 |
| CS02 | CX74-06257-1 *(§8.1 legend, p.67)* | 1 |
| CS03 | CX74-06258-1 *(§8.1 legend, p.67)* | 1 |
| CP01 | CX74-06278-1 *(§8.1 legend, p.67)* | 1 |
| HMS04 | n/a in source | 1, optional |
| Cable ties | n/a in source | as needed |
Safety
Hazard: Applying a supply voltage other than the one specified (100 VAC or 200 VAC) to a warning lamp or to its power-supply terminals, or using a fuse other than the specified 250 V, 0.5 A rating on the warning-lamp AC line
Consequence: The equipment may be damaged; equipment damage may also result from an abnormality if the wrong fuse is used
Avoidance: Never apply a supply voltage other than that specified to any warning lamp or to the power-supply terminals for lighting it; always use the specified 250 V, 0.5 A fuse
Full step-by-step for Connect warning lamp, door switch and inter-unit cables
All 20 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Power guides
Source: 2C201-160E*F p.73-80. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.