Connect warning lamp, door switch and inter-unit cables

Wire the warning lamp and door switch into TB2/TB3 with the correct fuse and JP1 jumper mode, then land the patient-couch, console (CPU BOX/REC BOX) and gantry-side inter-unit cables at their labeled connectors and clamp points, completing the electrical hookup of the Aquilion 16 before the pre-power checks begin.

Applies to
Aquilion 16 (TSX-101A) — new-installation site work, Ch.8 "Connecting Cables" (§8.3-8.7: warning-lamp connection methods, door-switch connection, the four patient-couch cables, console CPU BOX/REC BOX cable connections, and gantry-side cable connections/clamping), source volume 2C201-160
Personnel
1 technician · Basic-Intermediate — land labeled cables and connectors (S01-S03, P01-P04, CP01, CS01-CS03, FCP40/FCS46/FCS12/FCE01, HMS04) on the terminal-board and connector positions called out in the wiring diagrams, select one of three KGTSP-board JP1 jumper positions for the warning-lamp operating mode, and clamp cables with cable ties (taking care at the thickly-coated section of optical-fiber cables). No electrical measurement or adjustment is described in this section; the general terminal-board screwdriver technique and torque values for TB1/TB2/TB3 are covered in the preceding atom. (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
20 steps
Source
2C201-160E*F p.73-80

Tools & parts

ItemPart numberQty
250 V, 0.5 A fusen/a in source2
KGTSP PWB JP1 jumpern/a in source — full 4-position table (incl. pins 7-8, unused here) in [OPCONTA, GCIFA and KGTSP PWB settings and connectors](toshiba-ct-aquilion16-control-opconta-gcifa-kgtsp-pwb.md)1, set to one of three positions
Jig cover mounting bracketn/a in source1
FCP40 (POWER)CX77-96088-1 *(§8.1 cable-designator legend, p.67 — not restated on p.73-80)*1
FCS46 (SIGNAL)CX77-80261-2 *(§8.1 legend, p.67)*1
FCS12 (SIGNAL)CX77-96087-1 *(§8.1 legend, p.67)*1
FCE01 (EARTH)CX77-96089-1 *(§8.1 legend, p.67)*1
S01BSX70-0391 *(§8.1 legend, p.67)*1
S02CX70-02675-1 *(§8.1 legend, p.67)*1
S03CX70-02735-1 *(§8.1 legend, p.67)*1
P01CX70-03024-1 *(§8.1 legend, p.67)*1
P04CX70-02725-1 *(§8.1 legend, p.67)*1
P02PX70-02582-1 *(§8.1 legend, p.67)*1
P03PX70-02583-1 *(§8.1 legend, p.67)*1
E01PX70-02585-1 *(§8.1 legend, p.67)*1
CS01BSX74-1358-01 *(§8.1 legend, p.67)*1
CS02CX74-06257-1 *(§8.1 legend, p.67)*1
CS03CX74-06258-1 *(§8.1 legend, p.67)*1
CP01CX74-06278-1 *(§8.1 legend, p.67)*1
HMS04n/a in source1, optional
Cable tiesn/a in sourceas needed

Safety

NOTICE

Hazard: Applying a supply voltage other than the one specified (100 VAC or 200 VAC) to a warning lamp or to its power-supply terminals, or using a fuse other than the specified 250 V, 0.5 A rating on the warning-lamp AC line

Consequence: The equipment may be damaged; equipment damage may also result from an abnormality if the wrong fuse is used

Avoidance: Never apply a supply voltage other than that specified to any warning lamp or to the power-supply terminals for lighting it; always use the specified 250 V, 0.5 A fuse

Full step-by-step for Connect warning lamp, door switch and inter-unit cables

All 20 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Power guides

View all Aquilion 16 (TSX-101A) guides

Source: 2C201-160E*F p.73-80. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.