Power-supply mode setting and protective-timer checks

Test-trip the gantry's fault-current breaker (NFB2), choose and reconfirm which of the three power-supply modes the DAS timer runs the gantry rotation section in, and — for auto or synchronized-switching mode — confirm the off-delay timer drops rotation-section power exactly one hour after the console is turned off.

Applies to
Aquilion 16 (TSX-101A) — fault-current protective breaker (NFB2), gantry rotation-section power-supply mode (DAS timer), and off-delay timer, source volume 2C201-160 (Installation), §10.5-10.7
Personnel
1 technician · Basic — operate the breaker's built-in test button and the DAS timer's manual switch, frequency switch, and set indicators, then time the resulting behavior against a clock; no disassembly or calibration equipment is called for. (skill assessment derived from procedure content — not an OEM-stated skill level)
Steps
14 steps
Source
2C201-160E*F p.92-96

Tools & parts

ItemPart numberQty
Fault-current protective breaker (NFB2) test buttonn/a in source1 (built-in)
DAS timern/a in source1 (built-in)
Clock or watchn/a in source — not named as a tool in this section (derived — not in source)1

Full step-by-step for Power-supply mode setting and protective-timer checks

All 14 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure