Gantry/couch interlock test procedures
Step through the Aquilion 16's full HEIGHT/TILT/SLIDE interlock test set — one control at a time, from the couch/gantry operating panel, the Auto-set/Auto-home footswitch, and the console — confirming that each interlock stops the couch or gantry at its printed position with the correct panel-lamp response, so the installed system's interlock behavior matches what Appendix 28.7.1-28.7.3 documents it should be.
Applies to
Aquilion 16 (TSX-101A) — Installation volume (source 2C201-160), Appendix 28.7.4 ("Contents of test"), p.199-203. This test set exercises the interlock positions and operation-interlock table documented in Appendix 28.7.1-28.7.3 — see toshiba-ct-aquilion16-install-interlock-positions-and-list, Related — which source separates by control-unit variant (CBTB-016A/016B). Source does not tag any individual test in this appendix (28.7.4.1-28.7.4.5) to a specific couch/control-unit variant by name; confirm the fitted variant and its own position/table data before treating any stop-position value below as universal.
Personnel
1 technician · Intermediate — a long functional-verification sequence run entirely from the couch/gantry operating panel, the Auto-set/Auto-home footswitch, and the console: operate one control at a time, then read the resulting stop position (HEIGHT indicator, tilt angle) and panel lamp state against the printed tolerance for that sub-test. No electrical measurement, cover removal, or KGTSM terminal session is described in this section — but correctly recognizing a blinking indicator, a lamp that should go out vs. remain out, and an interlock-abort (operation must stop/not proceed) takes familiarity with the couch/gantry interlock system as a whole. (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
21 steps
Source
2C201-160E*F p.199-203
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Couch/gantry operating panel | n/a in source | |
| Auto-set/Auto-home footswitch | n/a in source | |
| Console | n/a in source | |
| External projector | n/a in source |
Full step-by-step for Gantry/couch interlock test procedures
All 21 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Couch/Table guides
Partially disassemble the patient couch for carrying-in (narrow route)5 stepsReset the overcurrent relay (horizontal and vertical motors)6 stepsFine-adjust the couch position using CT images13 stepsPrepare the couch and perform rough positioning with the projector16 stepsSecure the couch, install the footswitch, fit covers and set the height indicator18 stepsCarry the patient couch in tilted, on a triangular stand (narrow route or small elevator)11 steps