B6F10000 — Raw-data transmission error on the COM-PRE link
Canon Medical (Toshiba) CT error code, reported by the COM-PRE (PRE data link).
What B6F10000 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code B6F10000 indicates: Raw-data transmission error on the COM-PRE link. The code originates in the COM-PRE — PRE / scan data transfer section of the COM-PRE (PRE data link). Severity level 1 (Information / dialog).
Reported action: The raw scan data did not transfer cleanly. Confirm the study is complete before the patient leaves and repeat the acquisition if anything is missing. Errors that keep repeating on raw-data transfer belong to the COM-PRE link — cabling, connectors and the boards at each end are where service should start..
Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 8:Device COM-PRE table, where the OEM text reads "Error in Raw Data transmission."