B6300000 — Error in PRE — pre-processing fault on the COM-PRE path
Canon Medical (Toshiba) CT error code, reported by the COM-PRE (PRE data link).
What B6300000 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code B6300000 indicates: Error in PRE — pre-processing fault on the COM-PRE path. The code originates in the COM-PRE — PRE / scan data transfer section of the COM-PRE (PRE data link). Severity level 1 (Information / dialog).
Reported action: Repeat the scan and verify the resulting series is complete. If the PRE error comes back, write down which protocols and collimations trigger it; a fault that follows one acquisition mode points somewhere different than one that appears on every scan, and that is the first thing service will ask..
Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 8:Device COM-PRE table, where the OEM text reads "Error in PRE."
Full troubleshooting for B6300000
The complete diagnosis, cause tree, and step-by-step remedy for the COM-PRE — PRE / scan data transfer subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated COM-PRE — PRE / scan data transfer codes
Related service procedures
Source: errmsg.eng — 8:Device COM-PRE table. Reference information re-authored by FSELIB — not a copy of the OEM manual.