Information / dialogCOM-PRE — PRE / scan data transfer

B6610000PRE low-count error — fewer counts than expected (COM-PRE)

Canon Medical (Toshiba) CT error code, reported by the COM-PRE (PRE data link).

What B6610000 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code B6610000 indicates: PRE low-count error — fewer counts than expected (COM-PRE). The code originates in the COM-PRE — PRE / scan data transfer section of the COM-PRE (PRE data link). Severity level 1 (Information / dialog).

Reported action: Look at the affected images for excess noise before accepting the study. Check whether the low-count message is tied to low-dose/low-mA protocols or shows up regardless of technique, and repeat the scan. If it keeps returning, the acquisition chain feeding the PRE is what service should inspect..

Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 8:Device COM-PRE table, where the OEM text reads "Error in PRE ( Low count )."

Full troubleshooting for B6610000

The complete diagnosis, cause tree, and step-by-step remedy for the COM-PRE — PRE / scan data transfer subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related COM-PRE — PRE / scan data transfer codes

Related service procedures

Discuss B6610000 with other field engineers in the FSELIB forum

Source: errmsg.eng — 8:Device COM-PRE table. Reference information re-authored by FSELIB — not a copy of the OEM manual.