Alarm — high severityCOM-PRE — PRE / scan data transfer

B6310000GCIF detected a wrong transmitted channel number (COM-PRE)

Canon Medical (Toshiba) CT error code, reported by the COM-PRE (PRE data link).

What B6310000 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code B6310000 indicates: GCIF detected a wrong transmitted channel number (COM-PRE). The code originates in the COM-PRE — PRE / scan data transfer section of the COM-PRE (PRE data link). Severity level A1 (Alarm — high severity).

Reported action: The channel count the GCIF received did not match what was expected, so inspect the images of that scan for artifacts before using them. Repeat the acquisition; if the channel-number error repeats, the COM-PRE data path and its connections need to be checked by service rather than retried by the operator..

Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 8:Device COM-PRE table, where the OEM text reads "Error in transmitted channel number detected by GCIF.", and it is graded as a high-severity alarm.

Full troubleshooting for B6310000

The complete diagnosis, cause tree, and step-by-step remedy for the COM-PRE — PRE / scan data transfer subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related COM-PRE — PRE / scan data transfer codes

Related service procedures

Discuss B6310000 with other field engineers in the FSELIB forum

Source: errmsg.eng — 8:Device COM-PRE table. Reference information re-authored by FSELIB — not a copy of the OEM manual.