B6310000 — GCIF detected a wrong transmitted channel number (COM-PRE)
Canon Medical (Toshiba) CT error code, reported by the COM-PRE (PRE data link).
What B6310000 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code B6310000 indicates: GCIF detected a wrong transmitted channel number (COM-PRE). The code originates in the COM-PRE — PRE / scan data transfer section of the COM-PRE (PRE data link). Severity level A1 (Alarm — high severity).
Reported action: The channel count the GCIF received did not match what was expected, so inspect the images of that scan for artifacts before using them. Repeat the acquisition; if the channel-number error repeats, the COM-PRE data path and its connections need to be checked by service rather than retried by the operator..
Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 8:Device COM-PRE table, where the OEM text reads "Error in transmitted channel number detected by GCIF.", and it is graded as a high-severity alarm.
Full troubleshooting for B6310000
The complete diagnosis, cause tree, and step-by-step remedy for the COM-PRE — PRE / scan data transfer subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated COM-PRE — PRE / scan data transfer codes