B6320000 — GCIF detected a wrong transmitted projection-data number (COM-PRE)
Canon Medical (Toshiba) CT error code, reported by the COM-PRE (PRE data link).
What B6320000 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code B6320000 indicates: GCIF detected a wrong transmitted projection-data number (COM-PRE). The code originates in the COM-PRE — PRE / scan data transfer section of the COM-PRE (PRE data link). Severity level A1 (Alarm — high severity).
Reported action: The number of projections that reached the GCIF did not match the scan, so the reconstruction of that acquisition is suspect — review it, and repeat it if there is any doubt. Note whether it correlates with rotation speed or scan length. Recurrence means the COM-PRE link and its connections must be checked by service..
Reported on Aquilion 16, Aquilion RX/V4-62R004, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and V3.20R003. It comes from the 8:Device COM-PRE table, where the OEM text reads "Error in transmitted projection data number detected by GCIF.", and it is graded as a high-severity alarm.
Full troubleshooting for B6320000
The complete diagnosis, cause tree, and step-by-step remedy for the COM-PRE — PRE / scan data transfer subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.