Pular para o conteúdo principal
Alarm — high severityRDD (Raw Data Disk Array)

8002000BRDD HDC not responding

Canon Medical (Toshiba) CT error code, reported by the RDD (Raw Data Disk Array).

What 8002000B means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 8002000B indicates: RDD HDC not responding. The code originates in the RDD (Raw Data Disk Array) section of the RDD (Raw Data Disk Array). Severity level A1 (Alarm — high severity).

Reported action: The RDD's HDC stopped answering the console. Power-cycle the disk array, check its status LEDs and its fiber-channel connection, then run the console's hardware test for the RDD; if the HDC still does not respond, it needs board-level service..

Toshiba/Canon CT consoles on Aquilion RX/V4-62R004, V3.20R003, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and Aquilion 16. The OEM text reads "Error in RDD HDC (no response)." on all six platform columns.

Full troubleshooting for 8002000B

The complete diagnosis, cause tree, and step-by-step remedy for the RDD (Raw Data Disk Array) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RDD (Raw Data Disk Array) codes

Related service procedures

Discuss 8002000B with other field engineers in the FSELIB forum

Source: errmsg.eng — 40:Device RDD table. Reference information re-authored by FSELIB — not a copy of the OEM manual.