Pular para o conteúdo principal
Information / dialogRDD (Raw Data Disk Array)

80020004RDD command rejected — device not in ROM state

Canon Medical (Toshiba) CT error code, reported by the RDD (Raw Data Disk Array).

What 80020004 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code 80020004 indicates: RDD command rejected — device not in ROM state. The code originates in the RDD (Raw Data Disk Array) section of the RDD (Raw Data Disk Array). Severity level 1 (Information / dialog).

Reported action: The RDD (Raw Data Disk Array) rejected a command because it was not in the ROM state at that moment. Let the RDD operation in progress complete and retry; if the RDD keeps rejecting commands, power-cycle the disk array and check its status LEDs before escalating..

Toshiba/Canon CT consoles on Aquilion RX/V4-62R004, V3.20R003, Aquilion ONE/V4.2, Aquilion S16, Asteion Super4 and Aquilion 16. The OEM text reads "Command rejected ( non-ROM state now)." on all six platform columns, in the RDD (Raw Data Disk Array) message table.

Full troubleshooting for 80020004

The complete diagnosis, cause tree, and step-by-step remedy for the RDD (Raw Data Disk Array) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RDD (Raw Data Disk Array) codes

Related service procedures

Discuss 80020004 with other field engineers in the FSELIB forum

Source: errmsg.eng — 40:Device RDD table. Reference information re-authored by FSELIB — not a copy of the OEM manual.