Troubleshoot X-ray tube flow and thermal faults
Work the Aquilion 16's two X-ray tube cooling-interlock faults — **X-ray tube flow fault** and **X-ray tube thermal fault** — using the OEM's own two decision-tree flowcharts (source §8.2.6-8.2.7, p.83-84).
Applies to
Aquilion 16 (TSX-101A) — X-ray & HV volume, source 2D201-064, sections 8.2.6-8.2.7 (X-ray tube flow fault / X-ray tube thermal fault)
Personnel
1 technician · Intermediate — electrical fault isolation on the tube-cooling interlock loop: monitoring the 200 VAC input, continuity/voltage checks at terminal board TB5, fuse checks (F4/F5) on the power supply unit, and jumpering TB5 pins as a bypass test — escalating to replacing the oil cooler, the X-ray tube, or the XC PWB when a check does not clear the fault (skill assessment derived from procedure content — not an OEM-stated skill level)
Steps
0 steps
Source
2D201-064E*H p.83-84
Full step-by-step for Troubleshoot X-ray tube flow and thermal faults
All 0 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated X-ray & HV (Cooling) guides
Run the console malfunction-diagnosis program (hwtest)0 stepsRun the independent MHR PWB test (diagnostics) from the RTM terminal4 stepsRemove the host PC unit (PX74-07959 GR-1/GR-2)4 stepsRemove the host PC unit (PX74-08766 GR-1/GR-2)5 stepsRemove and install an RDD disk/power-supply unit (Rorke Data)5 stepsRemove and install an RDD fan unit (Rorke Data)5 steps