ProcedureX-ray & HV (X-ray tube)Functional ChecksMeasure
Check the rotor (starter) driver output
Confirm the SRU power supply unit's starter (rotor) driver is actually putting out an AC signal — a short, two-step voltmeter check at terminal stand TB5 that only judges whether voltage is present, not its exact value.
Applies to
Aquilion 16 (TSX-101A) — X-ray & HV volume 2D201-064, section 7.3 ("Checking the Rotor Driver")
Personnel
1 technician · Basic — inhibit gantry rotation, then read a voltmeter across two terminals of a labeled terminal stand and judge presence/absence of an AC signal; no calibration, waveform capture, or disassembly is asked for. (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
3 steps
Source
2D201-064E*H p.73
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Voltmeter | n/a in source |
Full step-by-step for Check the rotor (starter) driver output
All 3 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated X-ray & HV (X-ray tube) guides
Handle and replace the high-voltage cable (tube and SRU ends)36 stepsX-ray tube malfunctions troubleshooting and malfunction-code reference3 stepsChecks and adjustments when replacing the X-ray tube6 stepsRun the console malfunction-diagnosis program (hwtest)0 stepsRun the independent MHR PWB test (diagnostics) from the RTM terminal4 stepsRemove the host PC unit (PX74-07959 GR-1/GR-2)4 steps