Handle and replace the high-voltage cable (tube and SRU ends)

Disconnect and reconnect the high-voltage cable at the X-ray tube end and at the SRU end, using the correct cleaning, silicone-compound and torque steps for whichever of the three cable variants (BSX71-0782 / BSX71-0588-12, BSX71-0841E, or the 75 kV IEC standard product) is actually fitted, and replace the welded rubber at the SRU end when it is disturbed.

Applies to
Aquilion 16 (TSX-101A) — X-ray & HV volume, source 2D201-064 §6.4 ("Handling the High-Voltage Cable"), the X-ray-tube/SRU cable-service family within Chapter 6 — between the grouped floating-section/power-cable/gantry-rotation notices (§6.1-6.3, p.45-46) and the fuse-replacement/main-capacitor-discharge precautions that continue this chapter (p.61).
Personnel
1 technician · Advanced — contamination-free handling governs the whole family: clean the bushing, receptacle, plug and silicon plate with absolute alcohol until all metal powder/dust is removed, then apply the *correct* silicone compound for the specific cable variant being connected (silicone grease/oil at the tube end for BSX71-0782 / BSX71-0588-12; oil only, never silicone grease, on the 75 kV IEC plug and silicon plate) — source names the wrong compound explicitly as a withstand-voltage risk. The BSX71-0841E variant adds a timed hair-dryer heating step, a rubber-glove/drying-agent/oil-syringe kit check, a staged bolt-and-spring tightening sequence (two bolts first, then four with springs, then re-torque the first two), and a 2-4 mm clearance pass/fail gate with an explicit "replace the cable" consequence if it fails. The SRU-end connection and the 75 kV IEC plug both call for controlled, torque-driven tightening. The three tube-end/SRU-end variants are not interchangeable mid-job — the technician must identify which cable is actually fitted before starting. (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
36 steps
Source
2D201-064E*H p.47-60

Tools & parts

ItemPart numberQty
Silicone compoundAZP393-01qty not stated
Absolute alcoholn/a in sourceqty not stated
Toothbrushn/a in source1
Torque driverKANON 450QLK1
Hexagonal wrenchn/a in source1
Silicone oilBSX71-0945Eqty not stated
Acetone (or absolute alcohol)n/a in sourceqty not stated
Rubber glovessupplied with cable kit1 pair
Drying agent packetsupplied with cable kit1
Injection syringe (pre-filled with oil)supplied with cable kit1
Hair dryern/a in source1

Safety

NOTICE

Hazard: Rotating the cable, or removing the bushing suddenly, from the X-ray tube receptacle while disconnecting the high-voltage cable at the tube end

Consequence: May damage the pin section on the tip of the bushing

Avoidance: Do not rotate the cable or remove the bushing suddenly from the X-ray tube receptacle. (Source prints this as a CAUTION; FSELIB classifies it NOTICE per the Z535 consequence mapping, since the stated outcome is bushing-pin damage — equipment only, no personal-injury outcome is described.)

NOTICE

Hazard: Rotating the cable after inserting the bushing into the X-ray tube receptacle, while connecting the high-voltage cable at the tube end

Consequence: May damage the pin section on the tip of the bushing

Avoidance: Do not rotate the cable after inserting the bushing into the X-ray tube receptacle. (Source prints this as a CAUTION; FSELIB classifies it NOTICE for the same equipment-only reason as above.)

NOTICE

Hazard: Uneven bolt tightness when torquing the welded-rubber connection bolt at the SRU end

Consequence: The connector may be damaged

Avoidance: Tighten each bolt a little at a time until the specified torque (9.8 N·m) is reached. (Source prints this as a CAUTION; FSELIB classifies it NOTICE, since the stated outcome is connector damage — equipment only.)

NOTICE

Hazard: Disconnecting the 75 kV IEC-standard cable plug during maintenance or inspection without replacing the silicon plate with a new one, or using silicon plates/oil not manufactured by Claymount Corporation

Consequence: May result in system malfunction

Avoidance: Replace the silicon plate with a new one whenever the cable plug is disconnected for maintenance or inspection; use only silicon plates and silicon oil manufactured by Claymount Corporation. (Source prints this as a CAUTION; FSELIB classifies it NOTICE, since the stated outcome is system malfunction — equipment only.)

Full step-by-step for Handle and replace the high-voltage cable (tube and SRU ends)

All 36 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure