ProcedureSoftware / System SetupSetup & CalibrationConfigure
System Setup utility: model/tube/HFG selection at install and after hardware change
Reach the Setup main window through the Engineering-password-then-Service-Engineer's-password click path, then use it to set the System/Model Name, Hospital Name, Line Frequency and Mic Control at installation, and the Tube & HFG / Reconstruction Unit selections after a hardware change.
Applies to
Aquilion 16 (TSX-101A) — both system/model variants this Setup utility supports: the 16-slice system (Aquilion/16, system code TSX-101A/G) and the 8-slice system (Aquilion/8, system code TSX-101A/F), configured separately for the scan system console and the image processing (Sub) system console. Source volume 2D260-074E*B, p.77-80 (Chapter 6, "System Setup").
Personnel
Configuration/software — no board-level work or physical tooling; requires authorization for the Setup-level ("Service Engineer's") credential, one tier above the general Engineering password used elsewhere in this manual (e.g. to launch DCA). (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
6 steps
Source
2D260-074E*B p.77-80
Full step-by-step for System Setup utility: model/tube/HFG selection at install and after hardware change
All 6 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Software / System Setup guides
Run the console malfunction-diagnosis program (hwtest)0 stepsRun the independent MHR PWB test (diagnostics) from the RTM terminal4 stepsRemove the host PC unit (PX74-07959 GR-1/GR-2)4 stepsRemove the host PC unit (PX74-08766 GR-1/GR-2)5 stepsRemove and install an RDD disk/power-supply unit (Rorke Data)5 stepsRemove and install an RDD fan unit (Rorke Data)5 steps
View all Aquilion 16 (TSX-101A) guidesSource: 2D260-074E*B p.77-80. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.