Acquire NRA and air calibration data (setup, procedure, flow, deletion)

Acquire NRA (phantom-based) and air (no-phantom) calibration data for each required field-of-view, scan-mode, and tube-voltage combination from the console's Calibration data acquisition window, in the manual's own recommended order, then delete stored conditions as the 400-condition limit is approached.

Applies to
Aquilion 16 (TSX-101A, 8-slice and 16-slice) — Software volume 2D260-074, source section 3.4 "Calibration Data Acquisition (NRA/air calibration)" (p.45-54), the fourth and final section of this volume's Chapter 3, immediately following Alignment Workflow (§3.1, p.38), Front/Rear Alignment (§3.2, p.39-40) and Left/Right Alignment (§3.3, p.41-44) — see Related. Two window items in this section (the Focus pos display field, and the Delete button's 400-condition note) explicitly state they apply to "16-slice and 8-slice systems," confirming this section covers both configurations the volume serves. This section's own page footers are chapter-relative ("3-8" through "3-17"), mapped here to absolute manual pages 45-54 consistent with this pilot corpus's pagination; the footer for chapter-page "3-9" (absolute p.46) does not survive in this section's own text extraction, though the surrounding content and figure sequence remain unambiguous. Two sibling volumes point back at this same section instead of repeating it: the DAS volume's own §6.2.3 states verbatim "For details, refer to the software service manual" (2D201-065E*D, p.96); the Installation volume reproduces the identical workflow in full at its own §28.8 (2C201-160E*F, p.204-213) — cited here only as cross-reference, not as this atom's own source (see Related and Step Group 4 below).
Personnel
Intermediate — launch Calibration data acquisition from Utility > Maintenance, distinguish the NRA (phantom-based) and Air Calib (no-phantom) workflows, set Acq Mode/Tube Voltage/Scan Mode/FOV conditions, run and interpret the interactive phantom-positioning scan sequence, manage the 400-condition storage limit via Delete, and follow the recommended FOV acquisition order (LL>L>M>S>SS); assumes phantom handling and console navigation familiarity, no board-level work or disassembly. (Skill assessment derived from procedure content — not an OEM-stated skill level.)
Steps
26 steps
Source
2D260-074E*B p.45-54

Tools & parts

ItemPart numberQty
Water phantom, sized to match the FOV being calibrated (LL/L/M/S/SS)n/a in source1

Full step-by-step for Acquire NRA and air calibration data (setup, procedure, flow, deletion)

All 26 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Software / Calibration guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D260-074E*B p.45-54. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.