Data transfer unit operation and cleaning check (GCIFA PWB) — 5.1/5.2
Run the Aquilion 16's periodic Data Transfer Unit check on the GCIFA PWB — confirm its bank of red error LEDs stays off through a supervised local test rotation, then clean the panel's LEDs and photodiode hood — to close out periodic-inspection items 5.1 and 5.2.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Test rotor (60-second rotation) | n/a in source | |
| KGTSM maintenance panel | n/a in source |
Safety
Hazard: Leaving SW1-2 on the GCIFA PWB set to ON, or otherwise failing to return it to OFF, after completing the 5.1 operation check.
Consequence: Normal scan images cannot be obtained — an abnormal image results, or the console displays the message "Acquired data is abnormal."
Avoidance: After completing the check, always return SW1-2 on the GCIFA PWB to OFF before returning the system to clinical scanning.