Calibration data acquisition (NRA / air calibration)
Acquire NRA (phantom-based) and air (no-phantom) calibration data for every field-of-view, scan-mode and tube-voltage combination the site requires, from the console's Calibration data acquisition window, in the manual's own recommended order — the Installation volume's full copy of the same workflow that Section 23 of the installation flow, and the DAS and Software volumes, all point back to.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Water phantom, sized to match the FOV being calibrated (LL/L/M/S/SS) | n/a in source | 1 |
Safety
Hazard: Performing calibration data acquisition before waiting at least 3 hours after system power-ON for detector stabilization.
Consequence: The acquired images may contain ring artifacts.
Avoidance: Wait for at least 3 hours after the power is turned ON to allow stabilization of the detector before performing calibration data acquisition.
Full step-by-step for Calibration data acquisition (NRA / air calibration)
All 18 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Detector/DAS guides
Source: 2C201-160E*F p.204-213. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.