Replace the SRU (SRU-H / SRU-L / SRU-POWER)
Replace the SRU — the slip ring unit's SRU-H, SRU-L, or SRU-POWER assembly — by locking the gantry rotation, unbolting the assembly, carting it out on the SRU replacement jig, and fitting the new unit in reverse.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Standard tools | n/a in source | 1 set |
| Ratchet for hexagonal socket head bolt | n/a in source | 1 set (M12) |
| SRU replacement jig (CLX-21620-1) | CLX-21620-1 | 1 set |
| — Jig main body (PLATE, MTG) | WX77-86855-1 | 1 |
| — Jig mounting bolts, M6 | n/a in source | 5 |
| Stepladder | n/a in source | qty not stated |
Safety
Hazard: Removing a heavy component from the gantry rotation section without first engaging the mechanical brake (rotation lock pin)
Consequence: The gantry rotation section may rotate due to the weight imbalance created by the part's removal, possibly resulting in personal injury
Avoidance: Engage the mechanical brake (rotation lock pin) before removing any heavy component from the rotation section
Full step-by-step for Replace the SRU (SRU-H / SRU-L / SRU-POWER)
All 16 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry — Slip ring guides
Source: 2D201-061E*H p.88-89. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.