Replace the projector lamps (median, internal, side)
Swap the median-line, internal, or side projector (laser positioning lamp) on the Aquilion 16 gantry and reconnect it, then confirm the projected laser line still lands where it should.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Screwdriver (type not specified in source) | n/a in source | |
| Replacement projector — median line, internal, or side (as applicable) | n/a in source | |
| Mounting screw, M4 × 6 mm (as labeled in Figure 7.1-1 and Figure 7.1-2) | n/a in source |
Safety
Hazard: Fitting the median-line projector's mounting screw longer than 6 mm
Consequence: The screw may come into contact with parts inside the projector, resulting in damage to the projector
Avoidance: Use a screw with a length of 6 mm or less when installing the median-line projector (Figure 7.1-1)
Hazard: Fitting the internal projector's mounting screw longer than 8 mm
Consequence: The screw may come into contact with parts inside the projector, resulting in damage to the projector
Avoidance: Use a screw with a length of 8 mm or less when installing the internal projector (Figure 7.1-2)
Full step-by-step for Replace the projector lamps (median, internal, side)
All 21 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry — Optical / projector guides
Source: 2D201-061E*H p.66-68. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.