Adjust the projectors (inner, median, external)

Align the Aquilion 16 gantry's inner, median, and external projectors (the gantry positioning-light lamps) against a water-phantom target so their light beam patterns land exactly where the couch top and scan plane expect them.

Applies to
Aquilion 16 (TSX-101A) — Gantry mechanism, source volume 2D201-061
Personnel
1 technician · Intermediate — CT gantry optical-alignment work. The task calls for precise phantom setup (taping solder strips to specific reference lines), reading a scanogram image to confirm alignment, and iterative light-beam adjustment against printed patterns or couch-top reference marks, with a stated tolerance of 1 mm or less at one checkpoint. No torque value, electrical isolation, or power-down is called out in source for this task. (Skill-level assessment derived from procedure content — not an OEM-stated skill level.)
Steps
16 steps
Source
2D201-061E*H p.97-99

Tools & parts

ItemPart numberQty
Water phantom (with phantom holder)n/a in source
Adhesive tapen/a in source
Solder (strips)n/a in source
Paper marked with lines (per Figure 8.3-6)n/a in source
Scanogram acquisition (system function, not a hand tool)n/a in source
KGTS (couch shift control)n/a in source

Full step-by-step for Adjust the projectors (inner, median, external)

All 16 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Gantry — Optical / projector guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-061E*H p.97-99. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.