Configure the detector temperature controller after replacement (AXC004-01 / -02)

After installing an Aquilion 16 detector temperature controller — either the AXC004-01 (products up to June 2004) or the non-interchangeable AXC004-02 (products from July 2004) — step its front-panel keys through initial setting, a 37.0°C set point, an unattended ~30-minute PID auto-tuning cycle, and a protection lock, confirming the lock holds before returning the system to use.

Applies to
Aquilion 16 (TSX-101A) — data acquisition section volume 2D201-065, source subsection 6.2.5 "Setting after replacing the temperature controller," p.100-112. This section documents the front-panel SETUP sequence only — the physical removal/installation of the AXC004 temperature-controller unit itself is not covered within this page range (n/a in source); the setup below begins once a controller of either type is already installed. Two non-interchangeable part numbers apply, each with its own step order and key sequence (covered separately below): AXC004-01 (products manufactured up to June 2004; source's own parts diagram for this type is labeled "E5CN") and AXC004-02 (products manufactured from July 2004). Source states plainly that "The AXC004-01 is not compatible with the AXC004-02" — confirm which type is physically installed, by its external appearance/indicator, before following either half of this procedure.
Personnel
1 technician · Intermediate — identify which of the two non-interchangeable controller types (AXC004-01 vs AXC004-02) is installed by external appearance, then drive a multi-parameter front-panel key sequence (Level key, Mode key, UP key, DOWN key) through 4 setup steps per type — including an unattended ~30-minute PID auto-tuning wait and a numeric-range confirmation — before locking the settings with a protection code; no board-level rework or external test equipment is involved (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
15 steps
Source
2D201-065E*D p.100-112

Tools & parts

ItemPart numberQty
Temperature controller front panel (Level key, Mode key, UP key, DOWN key)AXC004-01 (labeled "E5CN" on source's parts diagram) or AXC004-02, whichever is already installed

Full step-by-step for Configure the detector temperature controller after replacement (AXC004-01 / -02)

All 15 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Detector temperature control guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-065E*D p.100-112. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.