Pular para o conteúdo principal
ProcedureCouch/Table (Covers)Replace

Remove and install the patient couch upper covers (handle, front, rear, side, rail, maintenance, couch-top bottom, and frame bottom covers)

Remove and reinstall the patient couch's upper cover set — handle, front cover, rear cover, side covers, rail covers, maintenance cover, couch-top bottom cover, and frame bottom cover — in the order that respects each cover's fit-up dependencies.

Applies to
Aquilion 16 (TSX-101A)
Personnel
1 technician · Basic-to-intermediate — hand-tool cover removal/installation sized to the M3/M4/M6/M12 fasteners called out in source (source names only thread size, not driver/wrench type); no electrical or hydraulic disconnection is described for these upper covers (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
9 steps
Source
2D201-062E*F p.22-29

Tools & parts

ItemPart numberQty
Handlen/a in source2 × M12 bolts
Front cover (lower)n/a in source5 × M6 screws + 2 × M3 screws
Front cover (upper)n/a in source3 × M4 screws
Rear covern/a in source4 × M4 screws (open the flap first)
Side cover \<L\>n/a in source15 × M4 screws
Side cover \<R\>n/a in source15 × M4 screws
Rail cover \<L\>n/a in source10 × M4 screws
Rail cover \<R\>n/a in source10 × M4 screws
Maintenance covern/a in source2 caps + 2 × M4 screws
Couch-top bottom covern/a in source8 × M4 screws
Frame bottom covern/a in source4 × M4 screws

Full step-by-step for Remove and install the patient couch upper covers (handle, front, rear, side, rail, maintenance, couch-top bottom, and frame bottom covers)

All 9 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Couch/Table (Covers) guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-062E*F p.22-29. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.