ProcedureCouch/Table (Covers)Replace

Replace the flexible (bellows) cover

Replace a damaged flexible (bellows) cover on the patient couch by working down from the bottom base covers, through the couch top and upper covers, to the vertical drive unit underneath, then reversing the sequence to fit the new cover.

Applies to
Aquilion 16 (TSX-101A)
Personnel
1 technician · Intermediate — sequential mechanical disassembly of couch covers and fasteners (M4/M8/M12), disconnecting cabling at the vertical drive unit, and physically supporting/handling the upper frame during removal; no electrical test equipment or calibration is called out in this section (skill assessment derived from procedure content — not an OEM-stated skill level)
Steps
8 steps
Source
2D201-062E*F p.42-47

Tools & parts

ItemPart numberQty
New flexible (bellows) covern/a in source1
Retaining plate A screwsn/a in sourceM4, 6 plates
Additional cover screwsn/a in source2-M4
Couch-top fixing boltsn/a in source2-M12
LM guide rail fixing boltsn/a in source15-M8
Upper-frame mounting boltsn/a in source11-M8
Square timber (or equivalent protective blocking)n/a in sourceAs needed

Safety

NOTICE

Hazard: The removed upper frame is set down without adequate protective blocking, or its cables and electric parts are damaged while it is handled (step 5).

Consequence: The PWBs and the horizontal drive unit carried on the upper frame are subjected to excessive force and can be damaged.

Avoidance: After removing the upper frame, place it on proper protective materials such as square timber, taking care not to damage the cables or electric parts, so the PWBs and the horizontal drive unit are not subjected to excessive force.

NOTICE

Hazard: During reassembly, the internal cable is routed so that it interferes with electromagnetic valve L24 (step 8).

Consequence: Source flags interference with electromagnetic valve L24 as a point requiring extreme care; the specific resulting fault is not detailed in this section.

Avoidance: While routing the internal cable, exercise extreme care to avoid interference with electromagnetic valve L24, following the routing detail in subsection 6.6.

Full step-by-step for Replace the flexible (bellows) cover

All 8 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Couch/Table (Covers) guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-062E*F p.42-47. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.