Use the KGTSM maintenance panel
Operate the Aquilion 16's KGTSM maintenance panel — or its PC-terminal equivalent, maintenance mode — to drive individual gantry, couch, slit/wedge, and scan functions directly for test and adjustment work, bypassing the normal operator panels and console.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| PC | n/a in source | |
| Service cable (GR-1) | GR-1 (standard accessory) | |
| Terminal program (e.g. Hyper Terminal) | n/a in source |
Safety
Hazard: Running the Scan function (SW32 = 8) more than 10 times in succession, or restarting the system sooner than 30 seconds after a scan shutdown.
Consequence: The controller for the DD motor may be damaged.
Avoidance: Do not perform scanning more than 10 times in succession while SW32 is set to 8. After any such shutdown, wait at least 30 seconds before restarting the system.
Full step-by-step for Use the KGTSM maintenance panel
All 18 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry-Couch/Control guides
Source: 2D201-063E*G p.54-59. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.