Pular para o conteúdo principal
ProcedureGantry-Couch/ControlOperate

Use the KGTSM maintenance panel

Operate the Aquilion 16's KGTSM maintenance panel — or its PC-terminal equivalent, maintenance mode — to drive individual gantry, couch, slit/wedge, and scan functions directly for test and adjustment work, bypassing the normal operator panels and console.

Applies to
Aquilion 16 (TSX-101A)
Personnel
1 technician · Intermediate — the operator must know in advance which SW32/SW31 combination is safe and appropriate to run (rotation, tilt, slide, scan, etc.) before pressing SET/SCAN, since several functions move the gantry, couch, or slit/wedge hardware directly (skill assessment derived from procedure content — not an OEM-stated skill level).
Steps
18 steps
Source
2D201-063E*G p.54-59

Tools & parts

ItemPart numberQty
PCn/a in source
Service cable (GR-1)GR-1 (standard accessory)
Terminal program (e.g. Hyper Terminal)n/a in source

Safety

NOTICE

Hazard: Running the Scan function (SW32 = 8) more than 10 times in succession, or restarting the system sooner than 30 seconds after a scan shutdown.

Consequence: The controller for the DD motor may be damaged.

Avoidance: Do not perform scanning more than 10 times in succession while SW32 is set to 8. After any such shutdown, wait at least 30 seconds before restarting the system.

Full step-by-step for Use the KGTSM maintenance panel

All 18 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.

Open the full procedure

Related Gantry-Couch/Control guides

View all Aquilion 16 (TSX-101A) guides

Source: 2D201-063E*G p.54-59. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.