AlarmRTM — ELCTL (pre-process / recon control)

C0900005No response from S&V reconstruction

Canon Medical (Toshiba) CT error code, reported by the RTM — pre-process / reconstruction control (ELCTL).

What C0900005 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code C0900005 indicates: No response from S&V reconstruction. The code originates in the RTM — ELCTL (pre-process / recon control) section of the RTM — pre-process / reconstruction control (ELCTL). Severity level A0 (Alarm).

Reported action: S&V reconstruction never answered. Check the reconstruction path (the MHR boards in the REC BOX) for a communication fault: review the event log, confirm the MHR boards are running, run the MHR hardware test, and retry the reconstruction..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003, Aquilion ONE (V4.2), Aquilion S16, Asteion Super4 and Aquilion 16 CT systems, reported under the RTM (real time manager) ELCTL device group of the system error message table. Reconstruction is performed by the MHR boards of the data processing unit in the REC BOX.

Full troubleshooting for C0900005

The complete diagnosis, cause tree, and step-by-step remedy for the RTM — ELCTL (pre-process / recon control) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.