C0900004 — No response to real-time reconstruction setup
Canon Medical (Toshiba) CT error code, reported by the RTM — pre-process / reconstruction control (ELCTL).
What C0900004 means
On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code C0900004 indicates: No response to real-time reconstruction setup. The code originates in the RTM — ELCTL (pre-process / recon control) section of the RTM — pre-process / reconstruction control (ELCTL). Severity level A0 (Alarm).
Reported action: The real-time reconstruction setup command went unanswered. Check the reconstruction path (the MHR boards in the REC BOX and the GLINK connections between them) for a communication or timing fault: review the event log, confirm the MHR boards are running, and run the MHR hardware test before rescanning..
Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003, Aquilion ONE (V4.2), Aquilion S16, Asteion Super4 and Aquilion 16 CT systems, reported under the RTM (real time manager) ELCTL device group of the system error message table. Reconstruction is performed by the MHR boards of the data processing unit in the REC BOX.
Full troubleshooting for C0900004
The complete diagnosis, cause tree, and step-by-step remedy for the RTM — ELCTL (pre-process / recon control) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated RTM — ELCTL (pre-process / recon control) codes
Related service procedures
Source: Toshiba/Canon CT service error message table (errmsg.eng, set 104 — Device RTM ELCTL): "Error in recon-process. (no responce to a real-time reconstruction setup.)". Reference information re-authored by FSELIB — not a copy of the OEM manual.