Pular para o conteúdo principal
AlarmRTM — ELCTL (pre-process / recon control)

C0900003No response when aborting pre-processing

Canon Medical (Toshiba) CT error code, reported by the RTM — pre-process / reconstruction control (ELCTL).

What C0900003 means

On Aquilion RX / V4-62R004 and V3.20R003 and Aquilion ONE / V4.2 and Aquilion S16 and Asteion Super4 and Aquilion 16 CT systems, status code C0900003 indicates: No response when aborting pre-processing. The code originates in the RTM — ELCTL (pre-process / recon control) section of the RTM — pre-process / reconstruction control (ELCTL). Severity level A0 (Alarm).

Reported action: An abort issued during pre-processing went unanswered. Check the data processing unit (the MHR boards in the REC BOX) for a communication fault: review the event log around the aborted scan, confirm the MHR boards are running, and retry the operation..

Toshiba/Canon Aquilion RX (V4-62R004), V3.20R003, Aquilion ONE (V4.2), Aquilion S16, Asteion Super4 and Aquilion 16 CT systems, reported under the RTM (real time manager) ELCTL device group of the system error message table. Pre-processing of the raw scan data is performed by the MHR boards of the data processing unit in the REC BOX.

Full troubleshooting for C0900003

The complete diagnosis, cause tree, and step-by-step remedy for the RTM — ELCTL (pre-process / recon control) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related RTM — ELCTL (pre-process / recon control) codes

Related service procedures

Discuss C0900003 with other field engineers in the FSELIB forum

Source: Toshiba/Canon CT service error message table (errmsg.eng, set 104 — Device RTM ELCTL): "Error in pre-process. (no responce to pre-process aborting)". Reference information re-authored by FSELIB — not a copy of the OEM manual.