ProcedureSoftware / AlignmentAdjust

Adjust front/rear X-ray tube alignment (Scan to calculate)

Launch the Aquilion 16's Alignment software in Front/Back mode and run "Scan to calculate" to measure and correct the X-ray tube's front/rear position, repeating the automatic scan-and-adjust cycle until the calculated movement distance falls within the 0.10 mm standard, per section 3.2 ("Front/Rear Tube Alignment") of the Aquilion16 PC Software service volume.

Applies to
Aquilion 16 (TSX-101A)
Personnel
1 technician · Adjustment/calibration — logging into the Engineering utility with a password, launching the Alignment software's Front/Back mode, running "Scan to calculate" (two automatic 0.5 s/rot scans), reading the calculated movement distance and direction, physically repositioning the X-ray tube to match, and repeating the scan/adjust cycle until the result is within the 0.10 mm standard. Source does not state a formal skill level (this assessment is derived from the procedure's content — not an OEM-stated skill level).
Steps
8 steps
Source
2D260-074E*B p.39-40

Tools & parts

ItemPart numberQty
Alignment software (Utility > Engineering login > Alignment, Front/Back mode)n/a in source1

Full step-by-step for Adjust front/rear X-ray tube alignment (Scan to calculate)

All 8 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.