Adjust the rotation balance (weights)
Adjust the rotation-section balance weights whenever any unit installed in the rotation section has been changed, by measuring the force needed to move the rotation section at four clock positions and correcting the front/rear weight distribution until the readings fall within tolerance of each other.
Tools & parts
| Item | Part number | Qty |
|---|---|---|
| Standard tools | n/a in source | 1 set |
| Push-pull gauge | n/a in source | 1 |
| Adjustment weights | see Table 8.9-1 | as needed, per Table 8.9-1 |
Safety
Hazard: Removing a part of the gantry rotation section without first engaging the rotation lock pin of the rotation drive section
Consequence: The rotation section may rotate suddenly due to the weight imbalance caused by removal of the part, and a worker may get caught in the rotation section and be seriously injured as a result
Avoidance: Engage the rotation lock pin of the rotation drive section before removing any part of the rotation section; hold the rotation section firmly when pulling out the rotation lock pin
Full step-by-step for Adjust the rotation balance (weights)
All 7 steps are documented in the FSELIB Aquilion 16 (TSX-101A) service manual — re-authored, web-first, with the real figures from the source.
Open the full procedureRelated Gantry — Rotation drive guides
Source: 2D201-061E*H p.116-118. Preview re-authored by FSELIB — the full procedure with figures and steps is in the paid service manual.